What is the difference between temperature sensor characteristic vs. sensor calibration - embedded

I'm configuring the internal temperature sensor of a STM32F4 MCU, and when reading the documentation, I came across some "duplicated" but divergent definitions.
Take a look in the image below:
The temperature sensor is connected to the ADC1, channel 16. When reading the ADC value inside my room, I always get values around ~920;
The values for the calibrations (read from MCU memory) are the following:
TS_CAL1 = 941
TS_CAL2 = 1199
It seems to me that calculating the final temperature using the relationship shown on Table 69 leads to different results from when using the relationship from Table 70.
Can anyone help me in this topic? What's the difference between the data of Tables 69 and 70? What is the purpose of each one? How to calculate the temperature correctly?

As Clifford explained in the comments, the information in table 69 tells you the typical behaviour of any device from this family, whereas the pointers in table 70 give you the address of the calibration data for your particular device which were measured in the factory.
If you told me that some device of this type gave a reading of 920, I would estimate the temperature as follows:
ADC voltage = 920/4096 * 3.3V = 741mV
Voltage offset from V(25C) = 741mV - 760mV = -19mV
Temperature offset from 25C = -19/2.5 = 7.6C
Temperature = 25 - 7.6 = 17.4 degrees C
For your calibrated device I would estimate the temperature like this:
Slope = (1199 - 941) / (110 - 30) = 3.225 LSB/degree
ADC offset from ADC(30C) = (920 - 941) = -21 LSB
Temperature offset from 30C = -21 / 3.225 = 17.775 C
Temperature = 30 - 17.775 = 12.2 degrees C
It is important to note however that although this second number is "calibrated", it is done so using calibration data from much higher temperatures. To use it below 30 degrees requires to extrapolate in a way which may not be physically valid.
Ask yourself, was the room closer to 17 degrees or 12 degrees? Bear in mind that the internal temperature sensor is probably subject to a certain amount of self-heating from the high performance processor.
If you want to use the internal temperature sensor to measure low temperatures outside the calibration range like this it might be appropriate to use the offset from the lower calibration point, but then use the typical slope from the datasheet.
Note also that many STM32 evaluation boards run at 3.0V not 3.3V, so all the calculation will have to be changed if that is the case.

I can not see the table but in theory there are following points need to considers when working with sensor:
Offset: value that is different with the actual value. You may check by feed a constant voltage to system and compare with ADC value measured.
Error of sensor: you might need to measure at known value. For example, for temperature, it is used to measure at 0 temperature which is a steady state.

Related

Adapting Smartphone Camera to derive Blackbody temperature

At first blush this presumably means -
(1) looking only at lower IR frequencies,
(2) select a IR frequency cut-off for low frequency buckets of the u/v FFT grid
(3) Once we have that, derive the power distribution - squares of amplitudes - for that IR range of frequency buckets the camera supports.
(4) Fit that distribution against the Rayleigh-Jones classical Black Box radiation formula:
(https://en.wikipedia.org/wiki/Rayleigh%E2%80%93Jeans_law#Other_forms_of_Rayleigh%E2%80%93Jeans_law)
(5) Assign a Temperature of 'best fit'.
The units for B(ν,T) are Power per unit frequency per unit surface area at equilibrium Temperature
Of course, this leaves many details out, such as (6) cancelling background, etc, but one could perhaps use the opposite facing camera to assist in that. Where buckets do not straddle the temperature of interest, (7) use a one-sided distribution to derive an inferred Gaussian curve to fit the Rayleigh-Jeans curve at that derived central frequency ν, for measured temperature T.
Finally (8) check if this procedure can consistently detect a high vs low surface temperature (9) check if it can consistently identify a 'fever' temperature (say, 101 Fahrenheit / 38 Celcius) pointing at a forehead.
If all that can be done, (10) Voila! a body fever detector
So those who are capable can fill us in on whether this is possible to do so for eventual posting at an app store as a free Covid19 safe body temperature app? I have a strong sense there's quite a few out there who can verify this in a week or two!
It appears that the analog signal assumed in (1) and (2) are not available in the Android digital Camera2 interface.
Android RAW image stream, that is uncompressed YUV, is already encoded Y green monochrome, and U,V are blue and red shifts from zero for converting green monochrome to color.
The original analog frequency / energy signal is not immediately accessible. So adaptation is not possible (yet).

Getting fuel% from analog data

I am getting analog voltage data, in mV, from a fuel gauge. The calibration readings were taken for every 10% change in the fuel gauge as mentioned below :
0% - 2000mV
10% - 2100mV
20% - 3200mV
30% - 3645mV
40% - 3755mV
50% - 3922mV
60% - 4300mV
70% - 4500mv
80% - 5210mV
90% - 5400mV
100% - 5800mV
The tank capacity is 45L.
Post calibration, I am getting reading from adc as let's say, 3000mV. How to calculate the exact % of fuel left in the tank?
If you plot the transfer function of ADC reading agaist the percentage tank contents you get a graph like this
There appears to be a fair degree of non linearity in the relationship between the sensor and the measured quantity. This could be down to a measurement error that was made while performing the calibration or it could be a true non linear relationship between the sensor reading and the tank contents. Using these results will give fairly inaccurate estimates of tank contents due to the non linearity of the transfer function.
If the relationship is linear or can be described by another mathematical relationship then you can perform an interpolation between known points using this mathematical relationship.
If the relationship is not linear than you will need many more known points in your calibration data so that the errors due to the interpolation between points is minimised.
The percentage value corresponding to the ADC reading can be approximated by finding the entries in the calibration above and below the reading that has been taken - for the ADC reading example in the question these would be the 10% and 20% values
Interpolation_Proportion = (ADC - ADC_Below) / (ADC_Above - ADC_Below) ;
Percent = Percent_Below + (Interpolation_Proportion * (Percent_Above - Percent_Below)) ;
.
Interpolation proportion = (3000-2100)/(3200-2100)
= 900/1100
= 0.82
Percent = 10 + (0.82 * (20 - 10)
= 10 + 8.2
= 18.2%
Capacity = 45 * 18.2 / 100
= 8.19 litres
When plotted it appears that the data id broadly linear, with some outliers. It is likely that this is experimental error or possibly influenced by confounding factors such as electrical noise or temperature variation, or even just the the liquid slopping around! Without details of how the data was gathered and how carefully, it is not possible to determine, but I would ask how many samples were taken per measurement, whether these are averaged or instantaneous and whether the results are exactly repeatable over more than one experiment?
Assuming the results are "indicative" only, then it is probably wisest from the data you do have to assume that the transfer function is linear, and to perform a linear regression from the scatter plot of your test data. That can be most done easily using any spreadsheet charting "trendline" function:
From your date the transfer function is:
Fuel% = (0.0262 x SensormV) - 54.5
So for your example 3000mV, Fuel% = (0.0262 x 3000) - 54.5 = 24.1%
For your 45L tank that equates to about 10.8 Litres.

Kinect Fusion volume voxel settings?

I need some help trying to figure out the Volume Voxel Per Meter and Resolution settings in Kinect Fusion...mostly how, and if at all, they interact with Depth Threshold settings in the Kinect Fusion Explorer program please...because I don't get if the depth threshold minimum is increased and maximum is reduced, does that smaller range increases the overall precision of the scanned volume, or does it stay the same?
Say I set the Kinect Fusion's depth threshold minimum to 2m and the maximum to 3m, thus setting the scanned range to 3m-2m=1m, does then the volume voxels per meter setting of say 256 and a resolution of also 256 mean that I would get a voxel depth precision of 1m/256=0.003m=0.3cm (a third of a centimeter)? Or is the resolution applicable only to the complete Kinect depth range instead of the one set via depth threshold? Also, how's width and height affected by depth threshold settings, and how to calculate precision in those two remaining axis?
Thanks in advance
P.S.
If the volume voxel resolution is set to maximum for all three axis (768x768x768) what is the minimum amount of GPU memory needed to make Kinect Fusion work?
Answering an old topic; because there is no other answer:
A. Simple Answer:
Depth threshold settings simply decide what region of the depth map are you interested in. Any value below min depth threshold and above max depth threshold is simply replaced with 0 during depth map generation.
B. Detailed Answer:
Volume Voxel per meter: This is the mm value depth represented by a single voxels . So 1000mm/256 (voxel_per_meter) = ~3.9 mm/voxel
( See:PCL documentation )
Voxel Resolution: The number of voxels in the volume you are constructing.
So;
Voxel Resolution / Voxel per m = Volume of Reconstruction volume (in meters)
EG: 512 voxels / 256 vpm = 2.0m (The volume of the reconstruction cube, given that the number of voxels per side of the cube are the same - each axis can be independently defined.)
If you have the Kinect SDK installed; see the descriptions of the following variables:
minDepthClip = FusionDepthProcessor.DefaultMinimumDepth;
maxDepthClip = FusionDepthProcessor.DefaultMaximumDepth;
voxelsPerMeter; voxelsX; voxelsY; voxelsZ;
So; these values are not dependent (or vice versa) on the depth threshold value.
A good example of using the depth threshold values is in the great video by Daniel Shiffman ([Kinect & Processing])

Calculating the maximum physical rate (Nyquist performance limitation) of an ADC onboard a microcontroller

I'm trying to evaluate the maximum physical rate (Nyquist performance limit) of the A/Ds integrated on board various PIC microcontrollers.
However, to do the calculation requires parameters that I'm not finding explicitly stated in the datasheets, specifically Tacq, Fosc, TAD, and divisor parameters.
I've proceeded by making some assumptions but would be helpful to have a sanity check -- am I doing the maximum physical rate calculations correctly?
For illustration purposes only, I've taken the simplest possible PIC10F220 that has an ADC. This is to focus specifically on the interpretation of Tacq, Fosc, TAD, and divisor parameters, and not to suggest that any practical functionality could be implemented on this very basic chip. (This is to Clifford's points in the comments below.)
Calculation:
Nyquist Performance Analysis of PIC10F220
- Runs at clock speed of 8MHz.
- Has an instruction cycle of 0.5us [4 clock steps per instruction]
So:
- Get Tacq = 6.06 us [acquisition time for ADC, assuming chip temp. = 50*C]
[from datasheet p34]
- Set Fosc := 8MHz [? should this be internal clock speed ?]
- Set divisor := 4 [? assuming this is 4 from 4 clock steps per CPU instruction ?]
- This gives TAD = 0.5us [TAD = 1/(Fosc/divisor) ]
- Get conversion time is 13*TAD [from datasheet p31]
- This gives conversion time 6.5 us
- So ADC duration is 12.56 us [? Tacq + 13*TAD]
Assuming 10 instructions for a simple load/store/threshold done in real-time before the next sample (this is just a stub -- the point is the rest of the calculation):
- This adds another 5 us [0.5 us per instruction]
- To give total ADC and handling time of 17.56 us [ 12.56us + 1us + 4us ]
- before the sampling loop repeats [? Again Tacq ? + 13*TAD + handling ]
- If this is correct, then the max sampling rate is 56.9 ksps [ 1/ total time ]
- So the Nyquist frequency for this sampling rate is 28 kHz. [1/2 sampling rate]
Which means the (theoretical) performance of this system --- chip's A/D with the hypothetical real-time handling code --- is for signals that are bandlimited to 28 kHz.
Is this a correct assignment / interpretation of the data sheet in obtaining Tacq, Fosc, TAD, and divisor parameters and using them to obtain the maximum physical rate, or Nyquist performance limit, of this chip?
Thanks,
You're not going to be able to do much processing in 8 instructions, but assuming you're just doing something simple like storing the incoming samples to a buffer, or detecting a threshold, then your analysis looks good.
The actual chips I'm considering for the design are the dsPIC33FJ128MC804 (with 16b A/D) or dsPIC30F3014 (with 12b A/D).
That is an important distinction; the dsPIC ADC supports ping-pong DMA transfers of multiple channels simultaneously, so can minimise the effective software overhead per sample. That makes the calculation a somewhat different one. You need to determine from the sample rate and the DMA buffer size the time between sample buffer interrupts; that is how much processing time you have to deal with each buffer. If you are using Microchip's DSP library, it gives precise cycle time formulae for each algorithm, and block processing is considerably more efficient that sample-by-sample processing.
My last project was on a dsPIC33 with two channels sampled at 48KHz and 32word sample buffers (giving 667us to process each pair of buffers). The software processing was therefore entirely independent of the sampling since by using DMA they take place simultaneously.

VB FFT - stuck understanding relationship of results to frequency

Trying to understand an fft (Fast Fourier Transform) routine I'm using (stealing)(recycling)
Input is an array of 512 data points which are a sample waveform.
Test data is generated into this array. fft transforms this array into frequency domain.
Trying to understand relationship between freq, period, sample rate and position in fft array. I'll illustrate with examples:
========================================
Sample rate is 1000 samples/s.
Generate a set of samples at 10Hz.
Input array has peak values at arr(28), arr(128), arr(228) ...
period = 100 sample points
peak value in fft array is at index 6 (excluding a huge value at 0)
========================================
Sample rate is 8000 samples/s
Generate set of samples at 440Hz
Input array peak values include arr(7), arr(25), arr(43), arr(61) ...
period = 18 sample points
peak value in fft array is at index 29 (excluding a huge value at 0)
========================================
How do I relate the index of the peak in the fft array to frequency ?
If you ignore the imaginary part, the frequency distribution is linear across bins:
Frequency#i = (Sampling rate/2)*(i/Nbins).
So for your first example, assumming you had 256 bins, the largest bin corresponds to a frequency of 1000/2 * 6/256 = 11.7 Hz.
Since your input was 10Hz, I'd guess that bin 5 (9.7Hz) also had a big component.
To get better accuracy, you need to take more samples, to get smaller bins.
Your second example gives 8000/2*29/256 = 453Hz. Again, close, but you need more bins.
Your resolution here is only 4000/256 = 15.6Hz.
It would be helpful if you were to provide your sample dataset.
My guess would be that you have what are called sampling artifacts. The strong signal at DC ( frequency 0 ) suggests that this is the case.
You should always ensure that the average value in your input data is zero - find the average and subtract it from each sample point before invoking the fft is good practice.
Along the same lines, you have to be careful about the sampling window artifact. It is important that the first and last data point are close to zero because otherwise the "step" from outside to inside the sampling window has the effect of injecting a whole lot of energy at different frequencies.
The bottom line is that doing an fft analysis requires more care than simply recycling a fft routine found somewhere.
Here are the first 100 sample points of a 10Hz signal as described in the question, massaged to avoid sampling artifacts
> sinx[1:100]
[1] 0.000000e+00 6.279052e-02 1.253332e-01 1.873813e-01 2.486899e-01 3.090170e-01 3.681246e-01 4.257793e-01 4.817537e-01 5.358268e-01
[11] 5.877853e-01 6.374240e-01 6.845471e-01 7.289686e-01 7.705132e-01 8.090170e-01 8.443279e-01 8.763067e-01 9.048271e-01 9.297765e-01
[21] 9.510565e-01 9.685832e-01 9.822873e-01 9.921147e-01 9.980267e-01 1.000000e+00 9.980267e-01 9.921147e-01 9.822873e-01 9.685832e-01
[31] 9.510565e-01 9.297765e-01 9.048271e-01 8.763067e-01 8.443279e-01 8.090170e-01 7.705132e-01 7.289686e-01 6.845471e-01 6.374240e-01
[41] 5.877853e-01 5.358268e-01 4.817537e-01 4.257793e-01 3.681246e-01 3.090170e-01 2.486899e-01 1.873813e-01 1.253332e-01 6.279052e-02
[51] -2.542075e-15 -6.279052e-02 -1.253332e-01 -1.873813e-01 -2.486899e-01 -3.090170e-01 -3.681246e-01 -4.257793e-01 -4.817537e-01 -5.358268e-01
[61] -5.877853e-01 -6.374240e-01 -6.845471e-01 -7.289686e-01 -7.705132e-01 -8.090170e-01 -8.443279e-01 -8.763067e-01 -9.048271e-01 -9.297765e-01
[71] -9.510565e-01 -9.685832e-01 -9.822873e-01 -9.921147e-01 -9.980267e-01 -1.000000e+00 -9.980267e-01 -9.921147e-01 -9.822873e-01 -9.685832e-01
[81] -9.510565e-01 -9.297765e-01 -9.048271e-01 -8.763067e-01 -8.443279e-01 -8.090170e-01 -7.705132e-01 -7.289686e-01 -6.845471e-01 -6.374240e-01
[91] -5.877853e-01 -5.358268e-01 -4.817537e-01 -4.257793e-01 -3.681246e-01 -3.090170e-01 -2.486899e-01 -1.873813e-01 -1.253332e-01 -6.279052e-02
And here is the resulting absolute values of the fft frequency domain
[1] 7.160038e-13 1.008741e-01 2.080408e-01 3.291725e-01 4.753899e-01 6.653660e-01 9.352601e-01 1.368212e+00 2.211653e+00 4.691243e+00 5.001674e+02
[12] 5.293086e+00 2.742218e+00 1.891330e+00 1.462830e+00 1.203175e+00 1.028079e+00 9.014559e-01 8.052577e-01 7.294489e-01
I'm a little rusty too on math and signal processing but with the additional info I can give it a shot.
If you want to know the signal energy per bin you need the magnitude of the complex output. So just looking at the real output is not enough. Even when the input is only real numbers. For every bin the magnitude of the output is sqrt(real^2 + imag^2), just like pythagoras :-)
bins 0 to 449 are positive frequencies from 0 Hz to 500 Hz. bins 500 to 1000 are negative frequencies and should be the same as the positive for a real signal. If you process one buffer every second frequencies and array indices line up nicely. So the peak at index 6 corresponds with 6Hz so that's a bit strange. This might be because you're only looking at the real output data and the real and imaginary data combine to give an expected peak at index 10. The frequencies should map linearly to the bins.
The peaks at 0 indicates a DC offset.
It's been some time since I've done FFT's but here's what I remember
FFT usually takes complex numbers as input and output. So I'm not really sure how the real and imaginary part of the input and output map to the arrays.
I don't really understand what you're doing. In the first example you say you process sample buffers at 10Hz for a sample rate of 1000 Hz? So you should have 10 buffers per second with 100 samples each. I don't get how your input array can be at least 228 samples long.
Usually the first half of the output buffer are frequency bins from 0 frequency (=dc offset) to 1/2 sample rate. and the 2nd half are negative frequencies. if your input is only real data with 0 for the imaginary signal positive and negative frequencies are the same. The relationship of real/imaginary signal on the output contains phase information from your input signal.
The frequency for bin i is i * (samplerate / n), where n is the number of samples in the FFT's input window.
If you're handling audio, since pitch is proportional to log of frequency, the pitch resolution of the bins increases as the frequency does -- it's hard to resolve low frequency signals accurately. To do so you need to use larger FFT windows, which reduces time resolution. There is a tradeoff of frequency against time resolution for a given sample rate.
You mention a bin with a large value at 0 -- this is the bin with frequency 0, i.e. the DC component. If this is large, then presumably your values are generally positive. Bin n/2 (in your case 256) is the Nyquist frequency, half the sample rate, which is the highest frequency that can be resolved in the sampled signal at this rate.
If the signal is real, then bins n/2+1 to n-1 will contain the complex conjugates of bins n/2-1 to 1, respectively. The DC value only appears once.
The samples are, as others have said, equally spaced in the frequency domain (not logarithmic).
For example 1, you should get this:
alt text http://home.comcast.net/~kootsoop/images/SINE1.jpg
For the other example you should get
alt text http://home.comcast.net/~kootsoop/images/SINE2.jpg
So your answers both appear to be correct regarding the peak location.
What I'm not getting is the large DC component. Are you sure you are generating a sine wave as the input? Does the input go negative? For a sinewave, the DC should be close to zero provided you get enough cycles.
Another avenue is to craft a Goertzel's Algorithm of each note center frequency you are looking for.
Once you get one implementation of the algorithm working you can make it such that it takes parameters to set it's center frequency. With that you could easily run 88 of them or what ever you need in a collection and scan for the peak value.
The Goertzel Algorithm is basically a single bin FFT. Using this method you can place your bins logarithmically as musical notes naturally go.
Some pseudo code from Wikipedia:
s_prev = 0
s_prev2 = 0
coeff = 2*cos(2*PI*normalized_frequency);
for each sample, x[n],
s = x[n] + coeff*s_prev - s_prev2;
s_prev2 = s_prev;
s_prev = s;
end
power = s_prev2*s_prev2 + s_prev*s_prev - coeff*s_prev2*s_prev;
The two variables representing the previous two samples are maintained for the next iteration. This can be then used in a streaming application. I thinks perhaps the power calculation should be inside the loop as well. (However it is not depicted as such in the Wiki article.)
In the tone detection case there would be 88 different coeficients, 88 pairs of previous samples and would result in 88 power output samples indicating the relative level in that frequency bin.
WaveyDavey says that he's capturing sound from a mic, thru the audio hardware of his computer, BUT that his results are not zero-centered. This sounds like a problem with the hardware. It SHOULD BE zero-centered.
When the room is quiet, the stream of values coming from the sound API should be very close to 0 amplitude, with slight +- variations for ambient noise. If a vibratory sound is present in the room (e.g. a piano, a flute, a voice) the data stream should show a fundamentally sinusoidal-based wave that goes both positive and negative, and averages near zero. If this is not the case, the system has some funk going on!
-Rick